Thermo Scientific Revco Laboratory Refrigerators and Freezers Installation and Operation
6 Cleaning
Wash the inside of the cabinet with a solution of one quart of
warm water and two tablespoons of baking soda.
Wash the outside with a mild detergent such as dishwashing
liquid. Do not use harsh or abrasive cleaners.
When the condenser tubing on the back is dusty, vacuum it clean.
CAUTION! Condensers should be cleaned at least
every six months. In heavy traffic areas, condensers
load with dirt more quickly. Failure to keep the
condenser clean can result in equipment warm-up or
erratic temperatures.
7 Troubleshooting
Troubleshooting Procedures
Problem Cause Solution
Unit does not
operate.
Power supply
1. Check that the cord is
securely plugged-in.
2. Plug another appliance into
the outlet to see if it is live.
3. If the outlet is dead, check
the circuit breaker or fuses.
Temperature
fluctuates.
Cold control
Make sure that the cold control
is set correctly.
Condenser
Make sure the condenser is
clean.
Unit warms up.
Door is open
Make sure the door is
completely closed.
Door seal
Check the door seal (see
Section 4.4).
Warm product
recently loaded
in unit
Allow ample time to recover
from loading warm product.
Power supply
Check for proper voltage to the
unit. If there is no voltage to the
unit, call an electrician.
Compressor
If the compressor is not
running, have an electrician
check for proper voltage to the
unit.
If the compressor is running,
contact an authorized service
provider or call the technical
support hot line for assistance.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific REVCO Series and is the answer not in the manual?

Thermo Scientific REVCO Series Specifications

General IconGeneral
DefrostManual
ControlMicroprocessor
Alarm SystemAudible and visual alarms for temperature deviations
MaterialStainless steel or coated steel exterior
ApplicationLaboratory, pharmaceutical
Voltage115V or 230V (depending on model)
ShelvesAdjustable shelves
Door LockYes
CertificationsUL
RefrigerantCFC-free