MaxQ 6000
Stackable Incubated and Refrigerated Shakers
Operating Manual and Parts List LT2120X1 (7004352) Rev. 11

Questions and Answers

  • C
    Charlene MilesAug 15, 2025
    How to fix a Thermo Scientific SHKE6000 (4359) Laboratory Equipment shaker that will not heat?
    • S
      Samantha HunterAug 15, 2025
      If your Thermo Scientific shaker isn't heating, make sure the 'TEMP ON' lamp is lit. If it's not, push the 'TEMP ON' key.
  • V
    Veronica RobertsAug 18, 2025
    What to do if my Thermo Scientific Laboratory Equipment shaker doesn’t operate?
    • S
      Susan FordAug 18, 2025
      If your Thermo Scientific shaker isn't operating, start by checking if the power cord is securely plugged in. If it is, verify that the power supply matches the requirements listed on the data label. Also, check the circuit breaker and fuse, resetting or replacing them if necessary. If the control panel lights are flashing, press any key. Finally, if the continual timer is flashing, reset it.
  • C
    Courtney GuerreroAug 21, 2025
    Why is the platform on my Thermo Scientific Laboratory Equipment not rotating correctly or has erratic speed?
    • A
      Angela TaylorAug 22, 2025
      If the platform on your Thermo Scientific Laboratory Equipment isn't rotating correctly or has an erratic speed, check the power to the motor and replace the motor if it's defective. Also, inspect the drive belt and replace it if it's worn, broken, or has slipped off the pulley.
  • M
    michael67Aug 25, 2025
    What does error code E08 mean on my Thermo Scientific Laboratory Equipment?
    • C
      crystaljenkinsAug 25, 2025
      Error code E08 on your Thermo Scientific Laboratory Equipment can indicate several issues. First, ensure the set point is not too close to the ambient temperature, adjusting it to at least 10 °C above ambient if you have an incubated model. Check the cooling system for proper operation. Inspect the connections on the PC1415X3 AC board. Consider replacing the PC1415X3 AC board if there's a failure in the cooling control circuit. Also, verify that the door is properly sealed by checking the rubber door gasket.
  • R
    Robert StewartAug 27, 2025
    What does E06 error mean on Thermo Scientific SHKE6000 (4359) and how to fix it?
    • J
      Jamie DouglasAug 28, 2025
      The E06 error on your Thermo Scientific Laboratory Equipment indicates that the OTP RTD temperature is above the OTP potentiometer setting. You can try increasing the OTP potentiometer setting or reducing the chamber set point. Also, ensure that the OTP RTD is properly connected. If the issue persists, there might be a failure in the heating circuit on the AC board, requiring replacement of the PC1415X3 AC board.
  • D
    Dennis SmithAug 30, 2025
    What to do if my Thermo Scientific SHKE6000 (4359) Laboratory Equipment displays ‘bAL’?
    • V
      Victoria RobinsonAug 30, 2025
      If your Thermo Scientific Laboratory Equipment displays 'bAL', it means the unit is vibrating, producing excessive noise, and is not leveled properly. Also, it can mean that the shaking platform load is too large, so reduce the load placed on the shaker platform.
  • J
    Jason WrightSep 2, 2025
    What does error code E02 mean on my Thermo Scientific SHKE6000 (4359)?
    • J
      jmitchellSep 2, 2025
      Error code E02 on your Thermo Scientific Laboratory Equipment indicates a failure in the element, which requires replacing the element. It could also mean the element SSR is defective, needing replacement. Check the connections on the PC1415X3 AC board to ensure they are properly connected. If the heating circuit on the AC board has failed, replace the PC1415X3 AC board.
  • N
    Natalie RodriguezSep 5, 2025
    What does error E12 on Thermo Scientific Laboratory Equipment mean?
    • T
      Tony GravesSep 5, 2025
      Error E12 on your Thermo Scientific Laboratory Equipment indicates a locked rotor condition. You should try to free the locked rotor. If that doesn't work, it could indicate a failure in the motor, requiring replacement. Alternatively, there might be a failure in the motor control circuit, necessitating replacement of the PC1415X1 control board.
  • R
    rebecca54Sep 8, 2025
    What does error code E01 mean on my Thermo Scientific Laboratory Equipment?
    • M
      mwhiteSep 8, 2025
      Error code E01 on your Thermo Scientific Laboratory Equipment indicates that the chamber RTDs are not connected, so ensure proper connection of the RTD. It could also mean the RTD is open, requiring replacement of TC1415X1. If neither of these resolves the issue, the control board may be defective, requiring replacement of the PC1415X1 control board.
  • T
    Tonya MckenzieSep 12, 2025
    What does error E13 mean on Thermo Scientific Laboratory Equipment and how can I resolve it?
    • A
      Angela MeyerSep 12, 2025
      Error E13 on your Thermo Scientific Laboratory Equipment can be caused by a shaker pan adjusted incorrectly, so readjust the 9 screws that secure the pan. Another cause can be obstruction between shaking platform and chamber wall, so remove any obstruction.

Summary

Safety Information

General Safety Warnings

Important operating and personal safety precautions, including electrical hazards.

Specifications

Features

Unpacking and Installation

Operation

Basic Operation Controls

Covers unit leveling, powering on/off, and setting shaking speed.

Temperature Control Features

Instructions for setting operating temperature and over-temperature protection.

Calibration Procedures

Steps for calibrating temperature and shaking speed for accuracy.

Maintenance

Troubleshooting

Error Code Reference

Explains various error codes related to heating, shaking, and general operation.

Thermo Scientific SHKE6000 (4359) Specifications

General IconGeneral
BrandThermo Scientific
ModelSHKE6000 (4359)
CategoryLaboratory Equipment
LanguageEnglish