27 Barnstead|Thermolyne • 800-553-0039
NOTE
If furnace is to be used regularly, the airline regulator may be left open to 20 psi.
Exhaust tubing hook-up.
NOTE
Appropriate exhaust must be provided to remove smoke and gases produced in an ashing
procedure.
Using accessory stainless steel tubing available from Barnstead/Thermolyne Corporation (part
number AY408X1A) or equal quality 2.5 inch I.D. stainless steel tubing, connect flexible tubing
from vent port at top of furnace case to an appropriate negative pressure exhaust system. This
exhaust system must be capable of handling smoke and gases produced in an ashing procedure.
NOTE
Failure to connect the exhaust port to an appropriate exhaust system will result in smoke and
gases filling the work area. Without the connection, gases and smoke will escape around the
door seal and at the rear of the furnace.
Furnace Programming
Refer to "Programmable Models" in this manual to program in the following values for performing ASTM
specification D3174.
PR1 (Ramp 1) value is 8°C/minutes
PL1 (Level 1) value is 500°C
PD1 (Dwell 1) value is 0 minutes
PR2 (Ramp 2) value is 6°C/minutes
PL2 (Level 2) value is 750°C
PD2 (Dwell 2) value is 120 minutes
PR3 (Ramp 3) value is END (for 8 ramp & 8 dwell controller)
SP (Setpoint) value is 20° (for 2 ramp & 2 dwell controller)
PLC (Loop Center) value is 1
Included with this furnace are two stainless steel trays which will hold crucibles of quartz jars. Use the
appropriate side for the type of crucible you are using. A removable handle is also provided to use in
loading and unloading the trays from the furnace.