During Operation
• Ensure that the operating air temperature is within the range -10°C to +40°C.
• Ensure that the air humidity does not rise above 90%.
Service Intervals
Changing of the pump chamber block should only be undertaken by trained
service personnel.
Work on electrical devices may only be carried out by skilled personnel in accordance
with the applicable engineering regulations
AP-30
AP-40
AP-60N
AP-60/80
AP-80H
AP-100 AP-120
Chamber block change:
(Diaphragm & chamber
Block are one set) 1)
max. 18.000 running hours or
latest after 36 months
Filter change/ cleaning Every 3 months
Recommended working
Pressure (rated pressure)
120 mbar 200 mbar 200 mbar 200 mbar
Operating pressure range 100…230 mbar 100…300 mbar 140…300 mbar 160…300 mbar
1) Based on operation at rated pressure +/-20%.