PTA522, PTA523 and PTA524 Automatic Silent Compressors Chapter 3: Operation
Page 6 ETN057391-D02
Figure 6 Filling the Oil
3.2. Electrical Connection
Warning
The compressor must be connected to an outlet protected by a suitable circuit breaker.
The cable powering the compressor or any extensions must have a wire cross section proportional to the length,
i.e. 1 mm
2
for lengths up to 3 m, and 1.5 mm
2
for lengths from 3 m to 20 m.
Install the compressor on a flat surface, in a dry, well ventilated room, where the temperature is not likely to rise above 35°C.
3.3. Pneumatic Connection
The pressurized outlet of the compressor unit is supplied with a push-fit connector suitable for 6 mm outside diameter
pressure hose. To fit the hose simply push in. To release the hose, push the blue collar towards the fitting, and pull the
hose in the opposite direction. See Figure 7.
Figure 7 (a,b) Fitting the Pneumatic Hose and (c) Removal of the Hose