6. ‘Zero weight’ will cease flashing, and ‘----‘ will be displayed for 4 seconds while
samples are obtained and averaged from load cell 1 with no load.
7. ‘Test weight’ will start flashing after the samples have been taken, and the default
test weight will be displayed. Use the up and down arrow keys to enter the
known weight on hand that will be used for calibration. Press OK once the
correct weight is displayed.
8. ‘rdY?’ will be displayed with ‘test weight’ still flashing. Place the weight on load
cell 1 so that it is balanced evenly in the middle. Press weigh or OK to continue
or the back key to go back to the last procedure.
9. ‘Test weight’ will cease flashing and ‘----‘ will be displayed for 4 seconds while
samples are taken from load cell 1 with the weight applied.
10. If there was only 1 load cell in the system go to step 16. Otherwise ‘bAr2’ will be
displayed and ‘zero weight’ will be flashing. Press tare, weigh or OK to
continue, or the back key to go back to the previous procedure.
11. ‘rdY?’ will be displayed and ‘zero weight’ will be flashing. Remove any weight
from load cell 2 (the load cell plugged into the right hand socket) and then press
tare, weigh or OK. Press back to go back to the previous procedure.
12. ‘Zero weight’ will cease flashing, and ‘----‘ will be displayed for 4 seconds while
samples are obtained and averaged from load cell 2 with no load.
13. ‘Test weight’ will start flashing after the samples have been taken, and the test
weight used for bar 1 will be displayed. Use the up and down arrow keys to
adjust the weight if necessary. Press OK once the correct weight is displayed.
14. ‘rdY?’ will be displayed with ‘test weight’ still flashing. Place the weight on load
cell 2 so that it is balanced evenly in the middle. Press weigh or OK to continue
or the back key to go back to the last procedure.
15. ‘Test weight’ will cease flashing and ‘----‘ will be displayed for 4 seconds while
samples are taken from load cell 2 with the weight applied.
16. ‘donE’ will be displayed. This is the final step. If OK is pressed at this point the
new settings will be committed to memory and a tare will be performed. Remove
the test weight before pressing OK.
The new settings will be remembered until another calibration procedure is performed.