SPECIFICATIONS:
SR-2 resistivity measurement range
o 4 PIN Method 0.1 Ω to 3.3 MΩ.
o 2 PIN Method 0.03 Ω to 10 MΩ
Accuracy of measurements of 10% or better.
Overload circuit that helps the user to identify the correct measurement range.
Measurement results displayed on a LCD display.
Unit operates from a 12V rechargeable battery.
Soil Resistivity meter can be recorded on DL-1 data logger.
Has Low Battery Voltage indicator.
LED charging light indicator.
ACCURACY:
The reading accuracy is ±10% on full range and ±5% on 85% of measurement range, which can be verified
by a decade box as required by ASTM G 57 95a Standard.
The instrument operates at 97Hz ± 10% AC current as defined by the ASTM G 57 95a standard to
prevent electrode polarization. The internal circuitry is designed in such way that it will reject any common
mode noise and external EMI interference.
EXTERNAL REFERENCES:
ASTM G57-95a Standard can be obtained by visiting www.astm.org.
AASHTO T-288 a Standard can be obtained by visiting www.transportation.org
NACE Internationals, “Peabody’s Control of Pipeline Corrosion, Second edition”, pages 84, 105
can be obtained by visiting www.nace.org