1-2
electric shock. When connecting a device to the instrument using the RS-232C
connector, contact Tomey Corporation or our local distributor.
– Ground the instrument appropriately. Otherwise you may get an electric shock.
– Do not hold the measuring head, chin rest, forehead pad, or joystick when
lifting the instrument. These components are detachable and the instrument
may drop, resulting in injuries.
– Install the instrument in a location not subject to direct sunlight, high
temperature and humidity, or air with significant dust, salt, and/or sulfur
content. These may cause failure or malfunction of the instrument.
– Install the instrument in a level and stable location free of vibration or
mechanical impact to ensure correct measurement, and prevent the instrument
from falling or being dropped, resulting in fire or fatal accidents.
– Install the instrument between the patient and physician so that they can face
with each other.
– Install the instrument in a location that it is sufficiently clear of any other
equipment that may interfere with the examination when using the instrument.
Precautions before using the instrument
– Check that all cables are connected correctly and completely.
– Check the sections that the patient will directly touch.
– Peel off the top sheet of chin rest paper and clean the forehead pad with a
cloth dampened with alcohol before measurement.
– Check that the instrument is correctly grounded.
– Check that the date set in the instrument conforms to the actual operation date
and time.
Precautions during operation
– Do not place any container with liquid in it on the instrument. Any liquid
entering the instrument may cause an electric shock or failure.
– If any smoke, offensive odor, or abnormal sound occurs, turn off the
instrument immediately, disconnect the power plug from the outlet, and contact
our local distributor or Tomey Corporation.