Devicestabilizationinprocess.
Battery is low and no more readings are possible.
Readingbeforedevicestabilization.
The ambient temperature is not within the range between 50°F ~104°F (10°C and 40°C).
Error 5~9, the system is not functioning properly.
Temperature taken is higher than 108.0°F (42.2°C).
Temperature taken is lower than 93.2°F (34°C).
Devicecannotbepoweredontothereadystage.
Solution:Waituntilstopsashing.
Solution: Replace the battery.
Solution:Waituntilstopsashing.
Solution: Allow the thermometer to rest in a room for at least 30 minutes at room temperature:
50°F ~104°F (10°C and 40°C).
Solution: Unload the battery, wait for 1 minute and then turn on.
Solution: Check the integrity of the hygiene cover and take a new temperature reading.
Solution: Make sure the hygiene cover is clean and take a new temperature reading.
Solution: Insert a new battery.
Clinical repeatability 0.18°C newborn to 1 year old, 0.17°C 1-5years old, 0.15°C 5years+
This product comes with a 12-month warranty.
ManufactureDateastheserialnumber(pleaseopenthebatterycover,itisshownontheinsideof
thedevice.)Ex.SN:E912A000001,therst“E”isExternal,thesecondnumber“9”isthe
manufactureyear2009,thethirdandthefourthnumber“12”isthemanufacturemonth,the
remainingguresaretheserialnumber.OurbatchcodeisXXXXVT.
BF type applied part.
TROUBLESHOOTINGTROUBLESHOOTING