When operating the autoclave, observe the environmental conditions given below.
<
Environmental Conditions
>
Attach the clear folder for storing the operation manual using the attached decorative screw.
An attachment hole is provided at each side.
※
After reading the operation manual, store it carefully in the clear folder.
④
Select a place which is not subject to direct sunlight and is
well ventilated. Install the autoclave with a clearance of
150 mm or more behind the autoclave and 50 mm or more
at both sides. Provide sufcient space at the front and
above the autoclave.
②
Place the attached caster stoppers under the casters to x
the autoclave at the place of installation.
※
Be sure to place the attached caster stoppers under all casters.
Otherwise, the chamber lid cannot be opened even though
the lid-lock releasing pedal has been pressed.
※
Do not place any objects other than the attached caster
stoppers under the casters.
③
: 10 to 35°C
: 860 to 1060 hPa
: 30 to 85
%
: 2°
Ambient temperature
Atmospheric pressure
Relative humidity
Maximum gradient