ー17ー
When operating the autoclave, observe the environmental conditions given below.
<
Environmental Conditions
>
Attach the clear folder for storing the operation manual using the attached decorative screw.
An attachment hole is provided at each side.
After reading the operation manual, store it carefully in the clear folder.
Select a place which is not subject to direct sunlight and is
well ventilated. Install the autoclave with a clearance of
150 mm or more behind the autoclave and 50 mm or more
at both sides. Provide sufcient space at the front and
above the autoclave.
5cm
5cm
15cm
Place the attached caster stoppers under the casters to x
the autoclave at the place of installation.
Be sure to place the attached caster stoppers under all casters.
Otherwise, the chamber lid cannot be opened even though
the lid-lock releasing pedal has been pressed.
Do not place any objects other than the attached caster
stoppers under the casters.
Caster
Caster stopper
: 10 to 35°C
: 860 to 1060 hPa
: 30 to 85
%
: 2°
Ambient temperature
Atmospheric pressure
Relative humidity
Maximum gradient