PERFORMANCE & SPECIFICATIONS
PERFORMANCE & SPECIFICATIONS
PERFORMANCE
• Switching the illumination level by three steps or switching to the infrared light
(Carry out infrared observation by DIGITAL CAMERA UNIT DC-4 (optional accessory) or by combin-
ing TV Relay Lens (optional accessory) or TV Attachment (optional accessory) with TV Camera with
Infrared Sensitivity.)
ELECTRIC RATING
Source voltage : DC 12V
Power input : 1.6W
DIMENSIONS AND WEIGHT
Dimensions : 51mm (W) × 31.6mm (D) × 271mm (H)
Weight : 50g
PURPOSE OF USE
This instrument is combined with the slit lamp and is used as an auxiliary illumination. It is also used
as the illumination for infrared observation of eyeball and its appendages.
OPERATION PRINCIPLE
This instrument is an auxiliary illumination whose light source is LED.
MAINTENANCE
REGULAR MAINTENANCE
Before starting operation, check the following points.
• The pins must be securely fitted into the guide hole of the slit lamp.
• The illumination light level can be changed by operating the lever. The illumination can be changed
to the infrared illumination light.
• The lever can be moved securely.