TRIOMED UNIVERSAL. USER MANUAL24
10.4. List of possible faults
and suggested remedies
Possible faults and suggested remedies are listed in Table 2.
Table 2
No
Signs of a fault
Likely reason
Remedy
1 the number of
the programme
is not shown on
the indicator of
the power unit
when the radia
-
tor is connected
The batteries are
defective or have
discharged.
replace the bat-
teries;
if following the
insertion of
non-defective
batteries the
device cannot
be switched on,
send it to be re
-
paired
2
the same prob
-
lem
there is no con
-
tact in the USB
connectors or
miniUSB in the
power unit or the
radiator
restore contract
in the connec
-
tors by connect-
ing the cable
again
3
the same prob
-
lem
the radiator is
defective
by connecting
other radiators
identify the de-
fective one and
send it to be re
-
paired
4
the same prob
-
lem
the connection
cable is defective
use a similar cable
to make sure that
the defect is in the
cable
Replace the ca-
ble with a similar
one.