6.1.3 DC Amps
Ranges: 1.5, 6, and 30 Amps DC.
Accuracy: ±1.5% of full scale 0 to 50 C.
±2.0% of full scale -40 to 85 C.
Burden Volts: 60 mV nominal at full scale. Leads add
34 milliohms typical resistance. Fuse in
1.5 Amp range adds 30 milliohms typical
resistance.
Overload: For the 1.5 Amp range, the fuse blows
before damage. The 6 Amp and 30 Amp
ranges are unfused but can withstand a
50 amp overload for 30 seconds.
6.1.4 Common DC Specifications
NMRR: >60 dB at 25 Hz, Code Function switch
in NORM position.
>35 dB at 60 Hz and >60 dB at 96 Hz,
Code Function switch in PEAK or
TIMING position.
Polarity Detect: All DC functions indicate upscale. A red
NEG indicator lights for inputs of nega-
tive polarity. For zero input, the NEG
indicator is off.