What to do if the BATTERY STATUS indicator flashes and there is no function when pressing a key on my Trumpf Medical Equipment?
P
Peter HernandezAug 5, 2025
If the BATTERY STATUS indicator on the column keypad of your Trumpf Medical Equipment flashes and there's no function when pressing a key, it means the operating table is discharged. To fix this, operate and charge the operating table using the mains power cable.
U
ukingAug 10, 2025
Why Trumpf TruSystem 7500 Medical Equipment operating table cannot be switched on?
T
Tony MendozaAug 10, 2025
If your Trumpf Medical Equipment operating table cannot be switched on, it may be discharged. Operate and charge the operating table with the mains power cable.
J
James FigueroaAug 15, 2025
Why (Reverse) Trendelenburg position cannot be fully attained on Trumpf TruSystem 7500 Medical Equipment?
S
Scott CervantesAug 15, 2025
If the (Reverse) Trendelenburg position cannot be fully attained on your Trumpf Medical Equipment, the tilt angle setting may be too high. Reduce the tilt to resolve this.
H
Heather StevensonAug 19, 2025
Why tilt position cannot be fully attained on Trumpf Medical Equipment?
T
Tonya GomezAug 19, 2025
If the tilt position cannot be fully attained on your Trumpf Medical Equipment, the (reverse) Trendelenburg angle setting may be too high. Try reducing the (reverse) Trendelenburg setting.
A
Antonio ThomasAug 25, 2025
Why leg sections cannot be adjusted far enough down on Trumpf Medical Equipment?
S
samuel60Aug 25, 2025
If the leg sections on your Trumpf Medical Equipment cannot be adjusted far enough down, the longitudinal travel might be too far toward the head end. Adjust the longitudinal travel toward the foot end.
B
Bradley JonesAug 29, 2025
Why back section cannot be adjusted far enough down on Trumpf Medical Equipment?
E
Elizabeth ReedAug 30, 2025
If the back section on your Trumpf Medical Equipment cannot be adjusted far enough down, the longitudinal travel might be too far toward the foot end. Adjust the longitudinal travel toward the head end.