2-6 Model 3786 Ultrafine Water-based Condensation Particle Counter
d. Remove the drain bottle connector.
e. Switch off the instrument and remove the power cord.
4. The UWCPC is ready for local transporting.
Figure 2-4
Drain Bottle Connection
Shipping and Storage
The following procedure must be used in cases where the UWCPC
must be shipped, stored, or subjected to freezing temperatures. The
instrument must be “dried” using the following procedure.
1. Disconnect any connections to the Sample Inlet.
2. Remove the source water bottle and dry completely.
3. Turn on the UWCPC and let it complete its warm-up cycle.