Firmware Commands B-9
SIT Set Injection Time
This command sets the water injection rate in seconds between injections. The factory
default setting is 12 seconds.
Command: SIT,xxxx Set Injection Time command
xxxx ranges 0 to 3600 seconds
Response: OK
Example: SIT,12 Command sets injection interval to 12 seconds
OK Response
SIT Command to read back the current setting
12 Response
SL Set Laser
The set laser command switches the laser on or off. The laser is always in the “on”
condition when the instrument is first powered up.
Command: SL,x Command to set laser on or off
x = 0 turns laser off, x = 1 turns laser on
Response: OK
Example: SL,0 Command turns laser off
OK Response
SL Command to read back the current setting
0 Response indicates laser is off
SLM Set Live-time Minimum
This command sets minimum amount of live-time in proportion to elapse time allowed
before the “OVER” annotation is shown on the primary display. The factory default
value is 400 which represents 40%.
Command: SLM,xxx Command to set live-time minimum
xxx = 0 to 1001
Response: OK Response
Example: SLM,400 Command to set live-time minimum to 40%
OK Response
SLM Command to read back the current setting
400 Response