EasyManua.ls Logo

TSI Instruments 3786 - Page 75

TSI Instruments 3786
83 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Firmware Commands B-17
Example D Record:
D,2,0,2.27e3,6.0,5.875,66784,0,308
S Record (status)
S, FL, AP,GT,OT,CT
S Status Record Identifier
FL Aerosol flow rate in cm
3
/min
AP Absolute Pressure in millibar
ST Saturator Temperature in degrees Celsius
GT Growth tube Temperature in degrees Celsius
OT Optics Temperature in degrees Celsius
Example S Record:
S,300,970,12.0,75.0,75.0
Z Record (scanning data)
Z,Mode,Flags,RN,N0,N1… ,N9
Z Record Identifier
Mode Current Operating mode (5 or 6)
Flags See status flags
RN Record number - (1 to 600)
Nx Equivalent number of particles counted in tenth second (0 to 99999999) based on live-
corrected counts or the photometric value.
Example Z record:
Z,5,0,12,0,0,0,12,182,518,641,896,887,871
A Record (raw analog -- diagnostic only)
A,a,b,c,d,e,f,g,h,i,j,k,l,m,n,o
A Record Identifier
a-o raw analog values (bits)
Example A record:
A,2382,3194,1,1108,4095,4095,2735,1284,518,641,896,2278,821,819,1958
C Record (flow and temperature control - diagnostic only)
C,a,b,c,d,e,f,g,h,i,j,k,l
C Record Identifier
a - l control bit values
Example C record:
C,2413,2422,2166,2276,2276,1468,821,820,2894,821,820,1011,2210,2222,2012

Related product manuals