Prepare the PortaCount Fit Tester for Operation 19
3. Open the alcohol fill
capsule by twisting the
storage cap off (counter-
clockwise).
4. Set the storage cap and
alcohol cartridge down on
a clean surface to prevent
contamination.
5. Open a 30-ml bottle of
alcohol. Invert the bottle
and insert the nozzle end
into the alcohol fill
capsule as far as
possible to make certain
alcohol is not
inadvertently sprayed
anywhere except into the
capsule.
6. Squeeze alcohol into the
alcohol fill capsule. When
30 ml is added to an
empty alcohol fill
capsule, the alcohol level
will rise to the fill-line
when the wick is
inserted.
7. Recap the empty alcohol
bottle immediately.