EasyManua.ls Logo

TSI Instruments PRESSURA RPM20 - Optional Relative Humidity Sensor Wiring to Model RPM20; Optional Nurses Station Communications Wiring to Model RPM10; RPM20

TSI Instruments PRESSURA RPM20
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
p. 13 of 16
Figure 19. Optional Relative Humidity Sensor Wiring to Model RPM20
NOTE: Model RPM10 does not support Relative Humidity Sensor.
NOTE: A 120Ω resistor should be installed on both ends of the RS-485 daisy-chain.
Figure 20. Optional Nurses Station Communications Wiring to Model RPM10 / RPM20