Page 52
12 Specification
12.1.1 Frequency Measurement
Frequency Span
1 MHz to 2700 MHz in one range - PSA2702
1 MHz to 1300 MHz in one range - PSA1302
Centre frequency plus Span, or Start frequency plus Stop frequency
2699 MHz (1 MHz to 2700 MHz) - PSA2702
1299 MHz (1 MHz to 1300 MHz) - PSA1302
270 kHz, or Zero Span with demodulation
Reference Frequency Accuracy for Start, Stop & Centre (Zero-Span)
frequencies
Reference Frequency Accuracy
Better than ± 10 ppm at 20
o
C
Better than ± 10 ppm over 10
o
C to 30
o
C
Better than ± 3 ppm per year
Phase Noise
Typically -90dBc/Hz at 100kHz offset at 500MHz
Resolution Bandwidth
Selectable as 1 MHz, 280 kHz or 15 kHz
Frequency Markers
0.1 kHz at all frequencies
1/270
th
of Frequency Span ± 0.1kHz + Reference Frequency Accuracy
Display of absolute and difference frequencies for both markers
12.1.2 Amplitude Measurement (Level)
Amplitude Range
dBm or dBµV (dB milliwatts or dB micro volts)
Selectable as -20 dBm or 0dBm (87 dBµV or 107 dBµV)
Amplitude Accuracy
Better than ± 1 dB at 50MHz at 10dB below ref. level (20
o
C ± 5
o
C)
Better than ± 1.5 dB relative to 50MHz over 1MHz to 2700MHz
Better than ± 1 dB over 50dB range down from reference level
Noise Floor
Better than -95 dBm average displayed noise floor for F > 3MHz
(reference level = –20 dBm, RBW = 15 kHz, video filter On)
Distortion and Spurii
< –60dBc for two signals(500MHz and 502MHz) at 10dB below
reference level
< –60dBc at 10dB below reference level (100MHz)
<– 55dBc, typically <– 60dBc