EN – 22
2.1.3 Repair
Always contact the manufacturer for repairs!
2.1.4 Disposal
The PABLO® system must not be disposed of in the same manner as normal waste, but
returned to the company TYROMOTION GmbH.
2.2 Monthly functionality check / Recurring check
2.2.1 Check list functionality test
The functionality checks described here must be performed on a monthly basis. Perform
the checks even if the PABLO® system indicates a malfunction (e.g. in case of unusual
sounds, elementary damages, etc.). The person responsible for checking the device must
be trained in handling and operating the PABLO® system.
Inspection:
Malfunction:
Resulting measure:
Protective covers
• Covers shake
• Covers missing
• Covers damaged
• Further use is prohibited
• Contact TYROMOTION GmbH
Externally visible deformations
• Parts bent out of shape
• Parts asymmetrical
• Parts defective
• Further use is prohibited
• Contact TYROMOTION GmbH
Cleaning
• PABLO® system
contaminated
• Further use is prohibited.
• Clean the contaminated parts
as described in chapter 4.5.3
Safety strap PABLO®
Handsensor
• Safety strap missing or
damaged
• Further use is prohibited
• Contact TYROMOTION GmbH
Click closure for connecting
PABLO® Motionsensor
• Connection PABLO®
Motionsensor does not
hold
• Further use is prohibited
• Contact TYROMOTION GmbH
Table 6: Inspection points
Immediately stop using the system if one of the described malfunctions in
Table 6 has occurred or is suspected.