EN-62
Technical information
EN-63
Technical information
Unit type: Cycle temperature computer for recording basal body
temperature
Method: Symptothermal method modied on the basis of the recog-
nised regulations of the WHO
Unit dimensions: 8 cm x 10.8 cm x 3.3 cm
Weight: approx. 110 g
Housing material: PVC-free, ABS-base plastic
Temperature sensor: PVC-free, TPE-base exible sensor
Measurement cable: Special cable with strain-relief device (kevlar)
Sensor: Thermistor
Customer service
The unit may only be repaired by the manufacturer or by an expressly
authorised organisation.
Please contact: UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Phone: +49 (0) 9342/924040
Fax: +49 (0) 9342/924080
E-mail: info@uebe.com
Internet: www.uebe.com
Technical data
Display: LCD display (liquid crystal display)
Display elements: Day, temperature value in °C, baby symbols, battery symbol,
time-axis display, fault indications, direction arrow
Temperature display: Absolute in °C (non-extrapolating)
Measurement range: 35.5 - 38.0 °C
Measuring accuracy: ±0.1 °C at ambient temperature 10 °C to 35 °C, rel. air humidity
up to 85 % (non-condensing)
Storage and trans-
port conditions:
Ambient temperature -10 to +50 °C, Rel. air humidity up to 85
% (non-condensing), Protect against moisture/humidity
Serial number: The unit has a serial number which is identied with SN:
please quote this number in the event of possible service
enquiries.
Battery: RENATA type CR2450N mercury-free button-cell battery for
approx. 500 measurements, battery to be replaced only by
UEBE Medical GmbH. The unit may only be repaired by the
manufacturer or by an expressly authorised organisation. Do
not open the unit, as this will invalidate the warranty.
Service: Should your unit show evidence of a fault, we ask you not to
attempt repairs yourself, but instead to send the unit with a
short description of the fault directly to UEBE Medical GmbH
for inspection.
Calibration: The unit does not have to be calibrated; the metrological accu-
racy of the unit will be automatically checked when the battery
is replaced by UEBE Medical GmbH.
Protection class IP: IP 20: Protected against solid foreign particles with a diameter
of more than 12.5 mm, no protection against water.
Additional equipment connected to medical electrical equipment must comply with
the respective IEC or ISO standards (e.g. IEC 60950 for data-processing equipment).
Furthermore, all congurations shall comply with the normative requirements for medical
systems (see IEC 60601-1-1 or Section 16 of the 3rd edition of IEC 60601-1, respec-
tively). Anybody connecting additional equipment to medical electrical equipment is a
system congurer and is therefore responsible for ensuring that the system complies with
the normative requirements for systems. Attention is drawn to the fact that local laws
take priority over the above-mentioned normative requirements. If in doubt, consult your
local specialised dealer or the Technical Service department.
If you are returning a unit from a non-EU country, please observe your
national customs regulations.

Table of Contents

Other manuals for uebe Cyclotest 2 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the uebe Cyclotest 2 plus and is the answer not in the manual?

uebe Cyclotest 2 plus Specifications

General IconGeneral
Branduebe
ModelCyclotest 2 plus
CategoryTest Equipment
LanguageEnglish