Customer service2.
The unit may only be repaired by the manufacturer or by an expressly
authorised organisation. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.: +49 (0) 9342/924040
Fax: +49 (0) 9342/924080
E-mail: info@uebe.com
Internet: www.uebe.com
Failure
encountered
Possible cause Corrective action
Date/time not
shown in display
Function accidentally
deactivated. Function was
not reactivated after bat-
tery replacement.
Activate date and time as
described on Page EN-57.
Technical data3.
Unit type: Digital automatic unit with electric pump for
measuring blood pressure at the wrist
Dimensions: approx. W = 70 mm x H = 70 mm x D = 80
mm with cuff
Weight: 121 g with cuff, not including batteries
Display: LCD display (liquid crystal display) for meas-
ured values and check displays
Memory: 60 measured values (stored automatically)
and average value (A)
Measurement
procedure:
Oscillometric determination of systole,
diastole and pulse
Reference procedure of
clinical testing:
Auscultatory measurement
Inflating pressure: Initially up to approx. 190 mmHg, then in
stages of 40 mmHg
Pressure display range: 0-300 mmHg