EasyManua.ls Logo

uebe visocor HM50 - Original Spare Parts and Accessories; Applicable Standards

uebe visocor HM50
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN-72
english
Technical information
E
EN-73
english
Technical information
E
Measurement range Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits: Blood pressure measurement: corresponds to
EN 1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5 %
Serial number: The unit bears a serial number which
provides clear identification.
Power supply: Battery type: 2 x 1.5 V, AAA, LR 03, durability:
over 700 measurements in 2 years
Cuff: Cuff type H5, wrist circumference 12.5 cm
- 21.5 cm
Operating conditions: Ambient temperature 10 to 40 °C, relative
humidity up to 85 %, non-condensing
Storage and transport
conditions:
Ambient temperature -20 to +50 °C, relative
humidity up to 85 %, non-condensing
Air deflation valve: Electronically controlled
Original spare parts and accessories4.
Subject to technical modifications.
Cuff type H5 (for wrist circumference 12.5-21.5 cm)
Part no. 2205001, PZN 9101582
Automatic switch-off approx. 3 minutes after end of measurement
IP code: IPX0 not protected
Applicable standards5.
EN 1060-1 : 1995 + A2 : 2009 Non-invasive sphygmomanometers
- Part 1: General requirements
EN 1060-3 : 1997 + A2 : 2009 Non-invasive sphygmomanometers
- Part 3: Supplementary requirements for electro-mechanical blood pres-
sure measuring systems

Table of Contents