7
1. Turn off the power and remove the battery case cover.
2. Remove and dispose of old 9V alkaline battery promptly in a
manner consistent with disposal regulations. Always use a 9V
alkaline battery. Ensure the polarity of the battery matches that
shown in the compartment.
3. Replace the battery case cover.
BATTERY REPLACEMENT
REPLACEMENT PARTS
OPTIONAL ACCESSORIES
Battery AB10
Replacement Sensor RS1000
Replacement Sensor Cap RS1010
Headset Earphones HE1000
Hard Carrying Case AC501
ELECTROMAGNETIC COMPATIBILITY
The European Council Directive 89/366/EEC requires that electronic
equipment does not generate electromagnetic disturbances that
exceed defined levels and has an adequate level of immunity to enable
it to be operated as intended. The specific standards applicable to this
product are detailed in the appendices.
Since there are many electrical products in use that pre-date this
directive and may emit electromagnetic radiation in excess of the
standards defined in the directive, it may be appropriate to check the
analyzer prior to use. The following procedure should be adopted:
1. Go through the normal start sequence in the location where the
equipment is to be used.
2. Switch on all localized electrical equipment that might cause
interference.
3. Check that all readings are as expected. (Generally, a low disturbance
is acceptable.) If not, adjust the position of the instrument to minimize
interference or switch off the offending equipment during the test.
There has been no awareness of any field based situation where
such interference has ever occurred since the writing of this manual
(April 1998). This advice is only given to satisfy the requirements of
the directive.
This product has been tested for compliance with the following generic
standards, and it is certified to be compliant.
EN 50081-1 • EN 50082-1
Specifications EC/EMC/K1/CD100A details the specific test configuration,
performance and conditions of use.