EasyManua.ls Logo

Ulthera UC-1 - Page 64

Default Icon
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 64 Ulthera, Inc
V
Code V
Transducer energy deliv ery
halted
Tap Scan Y to resume scanning
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
Excessiv e reflected power has been
detected. If the problem persists,
please try another transducer and
contact Ulthera Support. Use
Transducer only as instructed. Tap
Scan Y to resume scanning.
W
Code W
Unauthorized transducer.
Please replace the transducer
and continue.
Please contact your local
representativ e for further
assistance.
The transducer connected is not an
authorized transducer. Contact your
local representativ e for further
assistance.
X
Code X
Transducer cannot be read
Please remov e and reinsert the
transducer.
If the problem persists please
see the User’s Manual for further
information or contact Ulthera
Support.
The transducer cannot be read.
Remov e and reinsert the transducer.
Check that the transducer contact
area is clean.