Ulthera, Inc. Page 49
4 List of Entered Regions
Lists the regions that hav e been entered into the user region treatment
guideline. The highlighted region is populated into the Edit Region Fields to
allow for editing.
If using the Edit Guideline user region action button, changes will not be
v isible in the therapy screen until a system reboot is performed.
5 Field Controls
Controls (keyboard or menu of options) used for entering information into
fields. This section changes based on the field that is being edited.
6 Sav e Button Sav es the current User Region Guideline and returns to the startup screen.
7 Add Region Button Adds a new, unnamed region to the treatment guideline.
8 Delete Region Button
Deletes the highlighted region.
NOTE: This action cannot be undone. In the event that a region is
inadv ertantly deleted the region will need to be re-entered.
9 Close Button
Closes the User Region Guideline dialogue without sav ing any
changes that hav e been made.
7.3. Operating Instructions
7.3.1. Activ ate the control unit:
1. Plug the power cord on the back of the system into the wall socket.
2. Turn the main power switch to the ON position.
3. The power switch may be left in the ON position when the system is not in
use.
Note: Do NOT use the power switch to shut down the system.
4. Insert the Ulthera System Access Key into the USB port on the back of the
control unit.
The Ulthera System operates only with the authorized Access Key.
5. Press the green On/Off button on front of the control unit.
The system will perform a brief self-test. After passing the self-test, a “NO
KEY message will be displayed if the Ulthera System Access Key has not
yet been inserted; otherwise the starting screen will be displayed.
Warning: If the self-test screen displays any information messages, turn
the system off by pressing the green On/Off button and follow the
instructions in the “Troubleshooting” section.