OTHER
FEATURES
1.
Labels MANUAL
and
PRESET
on the LCD
display, will
switch
on
according
the
program;
2.
Labels ACUTE
and
CHRONIC will switch
on
according
to
the
Pulse
Width
(50 us =
Gate
Control,
then most
suitable
to
ACUTE pain
-
Above
50 us =
Endorphins
then
most
suitable
to
chronic
pain),
regardless
the PW
is
manually
or
automatically adjusted.
3.
Symbols
® ® ©
will show
the
o c c u r r i n g phase
( i f
blinking)
or the
passed
phase
(if
permanently
ON) of
the
PRESET programs.
4.
Labels CONT,
MOD and
BURST have
not to be
shown
during
the
PRESET
programs.
5.
When mode changes,
the
intensity will automatically
go
down
to
zero;
6.
When
the
unit
is
turned
on, it
will automatically enter
the
mode that
the
unit
had
worked
in
before
the
unit
was
turned off;
7.
When
the
treatment timer
is
set,
it
will begin
to
count
down
one
minute
by one
minute; Once
it
count down
to
zero,
the
unit will automatically shut off;
8.
When
the
pulse rate, pulse width
or
intensity
is
being
adjusted,
the
auto-repeat function
is
available, when
the key is
held
down over
1
second,
it
will
automatically
increased
or
decreased
1
step
per 1/4
second.
-17-
9. The
treatment time will
be
accumulative recorded when
the
output level
is
above zero,
by
pressing "Timer"
key
and
"PR-"
key for 4
seconds
the
treatment time
accumulated
in
minute
can be
displayed,
or
return
back
to the
previous normal display
in
toggle;
by
pressing
"Timer"
key and
"PW-"
key for 4
seconds will
clear
the
treatment time
to
zero.
10.
When
the
unit
intensity
levels
are 0 on
both
channels,
and it has not
been
in use for 5
minutes,
the
unit will
be
shut
off
automatically.
Detail
information
regarding
to the
Waveform,
Pulse
duration,
Pulse
frequency,
Output
voltage
range,
and
Peak
pulse
output
current:
A.
Waveform
There
are 3
types
of
waveforms:
(1).
Symmetrical Bi-Phasic
rectangular
waveform
Amplitude
>-Time
(2). Asymmetrical Bi-Phasic rectangular w a v e f o r m
Amplitude
-18-

Questions and Answers: