UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:SYSTem:MNUDisplay
Command format:
:SYSTem:MNUDisplay { 1S | 2S | 5S | 10S | 20S | INFinite}
:SYSTem:MNUDisplay?
Functional description:
Set menu display timeINFinite present menu stay on.
Return format:
Query return{ 1S | 2S | 5S | 10S | 20S | INFinite }.
For example:
:SYSTem:MNUDisplay 1S set menu display time as 1s, fold menu fold automatically
:SYSTem:MNUDisplay? query return 1S
:SYSTem:BRIGhtness
Command format:
:SYSTem:BRIGhtness <count>
:SYSTem:BRIGhtness?
Functional description:
Set screen luminance, <count> take value from 1~100the number greater the screen brighter .
Return format:
Query return the current screen luminance.
For example:
:SYSTem:BRIGhtness 50 set screen luminance as 50
:SYSTem:BRIGhtness? query return 50
:SYSTem:VERSion?
Command format:
:SYSTem:VERSion?
Return format:
Query return version information, character string information of 128 bytes.
HW is hardware version number, SW is software version numberPD is creation dateICV is
protocol version number.
For example:
:SYST:VERS? query returnHW:1.0;SW:1.0;PD:2014-11-20;ICV:1.4.0