EasyManua.ls Logo

UNI-T UPO3000E Series - Page 10

UNI-T UPO3000E Series
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:SYSTem:CYMOmeter
Command format:
:SYSTem:CYMOmeter {1 | ON} | {0 | OFF}
:SYSTem:CYMOmeter?
Functional description:
The switch of frequency meter.
Return format:
Query return frequency meter status1 presents on, 0 presents off.
For example:
:SYSTem:CYMOmeter ON turn on frequency meter
:SYSTem:CYMOmeter? query return1
:SYSTem:SQUare:SELect
Command format:
:SYSTem:SQUare:SELect { 10Hz | 100Hz | 1KHz | 10KHz }
:SYSTem:SQUare:SELect?
Functional description:
Select square wave output.
Return format:
Query return{ 10Hz | 100Hz | 1KHz | 10KHz }.
For example:
:SYSTem:SQUare:SELect 10Hz select 10Hz square wave output
:SYSTem:SQUare:SELect? query return 10Hz
:SYSTem:OUTPut:SELect
Command format:
:SYSTem:OUTPut:SELect { TRIGger | PASS_FAIL }
:SYSTem:OUTPut:SELect?
Functional description:
Set output selection, TRIGger, PASS_FAILpass &fail.
Return format:
Query return{ TRIGger | PASS_FAIL }.
For example:
:SYSTem:OUTPut:SELect TRIG select TRIG
:SYSTem:OUTPut:SELect? query return TRIG

Other manuals for UNI-T UPO3000E Series

Related product manuals