UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
Query return1.120e000,unit is V.
:MEASure:VRMS?
Command format:
:MEASure:VRMS? [<interval>][,<source>]
Functional description:
Measuring the root mean square value of the specified channel waveform, <source> take value as
CHANnel1, CHANnel2, CHANnel3, CHANnel4 and MATH. If there has no designated <source>,
the current channel as <source> by default; <interval> appoint the measurement separation, take
value as CYCLe and DISPlay, CYCLe presents integer cycle period, DISPlay presents full screen.
If there has no designated <interval>, take DISPlay by default.
Return format:
Query return1.230e000,unit is V.
:MEASure:AREa?
Command format:
:MEASure:AREa? [<interval>][,<source>]
Functional description:
Measuring the area of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3, CHANnel4 and MATH. If there has no designated <source>, the current
channel as <source> by default; <interval> appoint the measurement separation, take value as
CYCLe and DISPlay, CYCLe presents integer cycle period, DISPlay presents full screen. If there
has no designated <interval>, take DISPlay by default.
Return format:
Query return 3.456e002,unit is Vs.
:MEASure:OVERshoot?
Command format:
:MEASure:OVERshoot? [<source>]
Functional description:
Measuring overshoot of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return1.230e002,unit is V.
:MEASure:PREShoot?
Command format: