:MEASure:PREShoot? [<source>]
Functional description:
Measuring preshoot of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return1.230e-002unit is V.
:MEASure:FREQuency?
Command format:
:MEASure:FREQuency? [<source>]
Functional description:
Measuring frequency of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return 2.000e003unit is Hz.
:MEASure:RISetime?
Command format:
:MEASure:RISetime? [<source>]
Functional description:
Measuring rise time of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 CHANnel4. Ignore means the current channel.
Return format:
Query return 5.000e-005unit is s.
:MEASure:FALLtime?
Command format:
:MEASure:FALLtime? [<source>]
Functional description:
Measuring fall time of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and CHANnel4. Ignore means the current channel.
Return format:
Query return 5.000e-005unit is s.
:MEASure:PERiod?
Command format:
:MEASure:PERiod? [<source>]
Functional description: