EasyManua.ls Logo

UNI-T UT58A - Accuracy Specifications(1); A. DC Voltage

UNI-T UT58A
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Specifications
Maximum Voltage or Current between input Terminals and
Grounding: According to different functional input protection
value.
l
Fused Protection for
l AmA Input Terminal : CE Version: 0.5A, 250V fast
type, 5x20mm.
l Maximum Display : 1999,updates 2~3 times/
second.
l Range : Manual ranging.
l Polarity display : Automatically
l Overloading : Display “1”.
l Battery Deficiency : Display “ ”.
l Temperature: Operating: 0
o
C~40
o
C (32
o
F~104
o
F);
Storage : -10
o
C~50
o
C(14
o
F~122
o
F)
A. DC Voltage
200mV
2V
20V
200V
1000V
0.1mV
1mV
10mV
100mV
1V
Range Resolution
(0.5%+1)
Overload
Protection
Accuracy
250VAC
(0.8%+2)
1000V AC
Remarks:Input Impedance: approx.10M.
B. AC Voltage
2V
20V
200V
750V
1mV
10mV
100mV
1V
Range Resolution
(0.8%+3)
Overload
Protection
Accuracy
(1.2%+3)
1000V AC
Remarks:
l Input Impedance: approx.10M.
l Frequency response:
40Hz~1kHz< 500V;
40Hz~400Hz > 500V
The
500Hz reading is for reference
l Displays effective value of sine wave (mean value
response).
Accuracy Specifications(1)
Accuracy:
(a% reading + b digits),guarantee for 1 year.
Operating temperature:18
o
C~28
o
C.
Relative humidity: 75%RH.
Model UT58A/B/C: OPERATING MANUAL