3
Unpacking and Installation:
The centrifuge is packed in cushioning Styrofoam in a box for maximum protection
during shipping. Please examine the unit upon receiving and retain all packing until the
unit has been tested. The warranty excludes any damage caused by shipping. Inspect for
any possible shipping damage. If you note any damage, please contact your sales
representative. Please save and store the original packaging. If it becomes necessary to
return the instrument to us, we ask that it be returned in the original packaging.
Unpack the shipping carton carefully and verify that the shipment is complete. It should
include the following items:
PowerSpin C8604 centrifuge 1 unit
Power cord 1 pc.
Operation Manual 1 pc.
Wrench 1 pc.
Screwdriver 1 pc.
Follow simple procedures below to prepare the centrifuge for use:
1. Remove the centrifuge from its packaging and place it on a sturdy flat, dry surface.
Do not plug-in or turn on the centrifuge!
2. MANDATORY: Place the centrifuge gently on its side. Using the enclosed
wrench, remove the two hex-pins in the center of the base. These are installed to
protect the unit during shipping and must be removed prior to using the centrifuge.
Please save these pins along with the packaging. You will need to re-install them if
the centrifuge will be shipped in the future.
3. Locate the small hole on the upper right corner of the left side of the centrifuge.
This is the Emergency lid latch manual override designed to open the
electronically controlled lid latch in the event of a power loss. Using a small
screw driver (or a rod with 4.5mm or less in diameter); insert the tool into the
emergency lid-opening hole as showed below. Push the tool to release the safety
lock and open the lid.
Special packing cushion foam is packed inside the rotor chamber of the centrifuge
to protect the rotor and shields during shipping.
REMOVE THIS FOAM PRIOR TO OPERATION OF THE CENTRIFUGE.
Top lid
Emergency
Lid-opening Front
4. The centrifuge rotor is pre-installed. No rotor installation or calibration is required.
5. The power inlet is located in the back of the centrifuge. Plug in one end of the
power cord to the inlet and the other end to an 115V power outlet.
Left Side