Do you have a question about the Unico PowerSpin C8604 and is the answer not in the manual?
Brand | Unico |
---|---|
Model | PowerSpin C8604 |
Category | Laboratory Equipment |
Language | English |
UNICO's commitment to innovation and product design for medical and health care industries.
Centrifuge designed for demanding lab work in clinical, industrial, or biological labs for material separation.
Highlights include digital controls, safety lid latch, maintenance-free motor, and speed/time memory.
Details the high-impact plastic exterior, stainless steel chamber, rubber mounts, and included rotor with tube shields.
Instructions for examining the unit upon receiving, verifying shipment contents, and saving packaging.
Steps to prepare the centrifuge, including removing shipping pins and locating the emergency lid-opening hole.
Diagram and explanation of the centrifuge's control panel layout and buttons.
Guide on how to set and adjust the timer (1-60 min) and speed (1,000-3,400 RPM).
Steps to initiate a run using the Start button, emphasizing lid closure for safety.
Functionality of the Stop button for immediate halt and the Open button for lid release when stopped.
Procedure for manually opening the lid using a tool in the emergency override hole during power loss.
How to check the total number of usage cycles by holding the Stop button for 4 seconds.
Step-by-step guide from plugging in and turning on to loading samples and completing a run.
Emphasis on zero RPM lid latch and the importance of never attempting to open the lid while the rotor is spinning.
Crucial instructions for balancing loads opposite each other in the rotor to prevent damage and ensure stability.
Information on cleaning the exterior and rotor chamber with approved solutions, avoiding solvents and immersion.
Requirements for rotor/tube holder inspection (6 months) and centrifuge calibration (initial, post-repair, 6 months).
Precautions regarding water/liquid spills, cord condition, and handling potential biohazards from broken tubes.
Contacting UNICO customer service, RGA process, and voiding warranty by unauthorized repairs or modifications.
Troubleshooting steps for issues like centrifuge not starting, excessive noise, and indicator light failure.
Detailed procedure for safely replacing the fuse, including necessary tools and fuse rating (250V 4A).
Details on speed range (1,000-3,400 RPM), speed accuracy (+/- 25 RPM), and digital timer (1-60 Minutes).
Information on rotor type (4 Place Horizontal Spin), capacity (Four x 50ml), dimensions, and net weight (32 lbs).
Covers defects in materials and workmanship for one year from shipment, excluding damage from accident or misuse.
Excludes repairs by unauthorized parties and specifies conditions for warranty claims and remedies.
Solutions for initialization failure and lid unable to open, involving checking connections and components.
Troubleshooting for centrifuge not sitting flat, no power, touch button response, and display issues.
Addresses problems like no RPM/RCF display, lid showing ready but not opening, sample separation failure, and shaking.
Labeled diagram identifying key parts of the UNICO PowerSpin C8604 centrifuge.
Identification of display windows, buttons, and indicator lights on the front control panel.