Varian ELSD Service Manual
Replacing The Nebuliser Chamber
When required: Blocked or contaminated Nebuliser.
Tools required: Allen keys
Spanners
Parts required: PTFE Nebuliser Chamber (#PL0890-0490)
Ensure that the detector has cooled down before proceeding
Ensure that all gas and solvent connections are disconnected
prior to proceeding
Open the detector and remove the
front panel as described in
“Removing the front panel”(pg 15).
Remove the nebuliser heater
assembly as described in
“Replacing the nebuliser heater
assembly”(pg 22).
The nebuliser and heater assembly
can be removed as one section, as
shown in diagram