Installation and use
Installation and use
08
English
07
English
▶ Timing/Min: display the device how many minutes the device are
working.
Remark: One screen showing Timing min & Accumulated hours. The default
display is accumulated hours. When pushing Timing/Min button, it is shift to
Timing minutes. If there is no new instruction, it is automatically switched back to
the accumulative hours display after 10 seconds.
▶ Gear: Output oxygen flow gear, a total of five gears (the five gears have
the largest flow)
Gear setting
Single pulse oxygen
flow rate (ML)
Allowable error is±20%
1 2
3
4
5
20 25
30
40 50
▶ Power indicator: Indicates the current built-in battery power, calculated
using voltage value, considering the actual usage, this item is for reference only;
There are a total of four battery indicator lights on the operation panel:
100% indicator light is on ---- battery power≥90%;
75% indicator light is on ---- 90%>battery power≧75%;
50% indicator light is on ---- 75%>battery power≧50%;
25% indicator light is on ---- 50%>battery power≧25%;
The 25% indicator flashes ---- battery power<25%.
▶ Indicator light and Alarm light:
Alam light:The light is on when the device is operated abnormally, meanwhile
will display the alarm code.
F0001 ---- Low Battery Warning;
F0002 ---- No Breath Detected;
F0011 ---- High Temperature with Alarm;
F0012 ---- Oxygen Sensor Abnormal Alarm;
Low Concentration:When the oxygen concentration is lower than 82%,the light
is on. When the oxygen concentration is lower than 50%, the light is on with
alarm.
Breath Signal: When the respiratory signal of the user is detected during the
operation of the equipment, the light is on. Meanwhile, the device will
automatically pulse oxygen supply.
3. Use Guidance