Installation Guide Resetting Data in the TLS RF Unit
34
TOTL COMMS
A running total of all messages received from this device. This includes messages that were repeated by a
repeater(s).
REPT COMMS Column
This column displays a running total of all messages received from this device that were repeated from a
repeater. Subtracting the repeated total from the total comms value will result in the total number of direct
messages received: Total Direct Messages = TOTL COMMS - REPT COMMS.
LAST COMM Column
This column displays the duration since the last message received from this device. Display is in
days:hours:minutes:seconds format. The 9999:99:99:99 time indicates no messages have been received
since the TLS RF unit has been powered on.
Resetting Data in the TLS RF Unit
NOTE: The data is reset when the TLS RF unit is power cycled or reset with the serial command S001
(except Constants which are in stored in flash memory).
It is also possible to reset the Total and Repeater Comms totals remotely with the following command:
S315ss149.

Other manuals for Veeder-Root TLS-350R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Veeder-Root TLS-350R and is the answer not in the manual?

Veeder-Root TLS-350R Specifications

General IconGeneral
BrandVeeder-Root
ModelTLS-350R
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals