DVM898
V. 01 06/04/2017 10 ©Velleman nv
6 A
1 mA
± (2.0 % rdg + 3 digits)
15 A
10 mA
Overload protection: F0.5 A/600 V fuse, F15 A/600 V fuse
Continuous testing max. steady-state current of 1 A
9.4 AC CURRENT
Do not measure circuits that may contain voltages
> 600 VDC or > 600 VAC
range
resolution
accuracy
600 µA
0.1 µA
± (1.5 % rdg + 3 digits)
6000 µA
1 µA
60 mA
10 µA
600 mA
100 µA
6 A
1 mA
± (2.5 % rdg + 5 digits)
15 A
10 mA
AC true rms
Frequency range: 40 Hz-1 kHz
Overload protection: F0.8 A/600 V fuse, F15 A/600 V fuse
Continuous testing max. steady-state current of 1 A
9.5 RESISTANCE
Do not conduct resistance measurements on live
circuits
range
resolution
accuracy
600
0.1
± (1.0 % rdg + 3 digits)
6 kΩ
1
± (1.0 % rdg + 2 digits)
60 kΩ
10
600 kΩ
100
6 MΩ
1 kΩ
60 MΩ
10 kΩ
± (1.5 % rdg + 3 digits)
Overload protection: 600 V DC or AC rms