What causes a high breath rate alarm on Ventec Life Systems Medical Equipment?
- LLisa EspinozaAug 28, 2025
If the High Breath Rate alarm activates on your Ventec Life Systems Medical Equipment, ensure the High Breath Rate alarm is set above the patient's breath rate plus spontaneous rate. Check for patient secretions and use Suction therapy to clear them if needed. Empty any excess condensation in the patient circuit. Check the VOCSN Leak monitor and resolve any unintentional leaks. If the device is auto-triggering, adjust the Flow Trigger setting.