Page 29BVI 3000 Operator’s Manual
Parts and Accessories
Part Number Description
0570-0090 BladderScan
®
BVI 3000 Instrument
0570-0091 Dome Scanhead Assembly
0400-0036 Battery Charger
0400-0039 Battery, 7.2 V
0800-0166 Carrying Case
0800-0167 Rolling Cart
0800-0005 Acoustic Coupling Gel, 0.25 liter
0800-0042 Printer Paper (5 rolls)
0900-0383 Operator’s Manual
0900-0396 Training Video
Technical Description
Compliance With
Standards
Verathon
®
certies that the BladderScan
®
BVI 3000 is in compliance
with all applicable international and national standards and regulations,
including, but not limited to, IEC-601-1 (EN60601-1) and EN 60601-1-
2; MDD 93/42/EEC Annex 1, CSA22.2 No. 125 (Risk class 2) (LR
76129-1), and UL544. Per the provisions of IEC-601-1, this device is
classied as:
Protection Class II, INTERNALLY POWERED
EQUIPMENT
TYPE BF, suitable for CONTINUOUS USE
Per the MDD, the BladderScan BVI 3000 is a Class IIa device.
Scanning
Specications
Temporal Average Power: 1 mW maximum
Focal 20 dB Beam-Area: 1.4 cm
2
Transducer Dimension: 3 mm diameter
Working Frequency: 2 MHz
Peak Instantaneous Intensity: 14 W/cm
2
maximum
Pulse Repetition Frequency: 180 pulses/second
Technical performance assessment (for example, in an acoustic
laboratory) may require partial immersion of the scanhead. Immerse the
black plastic dome no further than the point at which it connects to the
black body.
Scanning Specications