EasyManuals Logo

Verathon BladderScan BVI 3000 User Manual

Verathon BladderScan BVI 3000
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
Page 32 BVI 3000 Operators Manual
Input Current
0.39 A/North American (LZA-) units. 0.25/
European (LZE-) units. 0.38 A/Universal (LZU-)
units
Output
No load to full load at rated voltage. Refer to unit
label
Input Connection
The power supply employs a direct plug in AC
prongs for wall plug in units
Insulation
The power supply is Class I with basic insulation
to each terminal
Technical
Specications
(continued)
Technical Description: Battery Charger
Indoor use
Temperature
50° - 104° F (10° - 40° C)
Transient
overvoltage
Category II
Pollution degree
II
Atmospheric
pressure
70 hPa to 106 hPa
Relative humidity
30% to 75%, non-condensed
Ventilation
The power supply is designed to operate with free
air convection.
Earth Terminal
The earth terminal provided in this unit is a
functional earth terminal, as indicated by the
ground symbol on the unit. It is not utilized in the
power supply to provide earth protection. Various
ground options are available upon request.
Fuses
The fuses used in the power supply are rated at
250 VAC, 2 A, quick acting.
Cleaning
No cleaning is needed. If desired, a dry or slightly
damp cloth may be used to clean the surface of
the power supply while it is disconnected from
the AC source. In no instance should water be
allowed to enter the unit
Environmental
Conditions For Use

Other manuals for Verathon BladderScan BVI 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Verathon BladderScan BVI 3000 and is the answer not in the manual?

Verathon BladderScan BVI 3000 Specifications

General IconGeneral
BrandVerathon
ModelBladderScan BVI 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals