EasyManua.ls Logo

Vernacare Compact - Page 40

Vernacare Compact
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vernacare Compact Technical Manual-Revision 3-Issued by N.Shields 11.2018
Page | 40
FAULT
EXPLANATION
OPERATOR CORRECTIVE
ACTIONS
MAINTENANCE CORRECTIVE
ACTIONS
‘LID SEALError
Message
Displayed
The lid seal has
not inflated
properly
Call maintenance
Check pressure sensor for correct
operation.
Check airline connections for leaks
Check Solenoid valve for correct
operation.
Check lid seal integrity
‘LID ERROR
CALL
ENGINEER’
Error Message
Displayed
The lid cannot
open or close
Call maintenance
Check lid open and lid closed sensors for
correct operation.
Hopper Base
Not Clean At
End of Cycle
Overloading the
machine
Running a cycle with the
machine empty will remove
any residue
N/A
Hopper Sides/
Lid Not Clean At
End of Cycle
The lid spray is not
working correctly
Call maintenance
Remove and clean the lid spray
Water Leaking
From Machine
N/A
Call maintenance
Water leaking from the lid Check vent
pipe for blockage, check pressure
switches operating correctly, check for
debris on seal, check seal for damage
Water leaking from tank Check level
sensor for correct operation, check
solenoid valve for leaks
Water leaking from pipework, check the
integrity of the tubing & hose clips.
Water leaking from flange, check the
mechanical seal

Other manuals for Vernacare Compact

Related product manuals