Page | 11
2.3 Electrical
2.3.1 Single-phase electrical system (230v, 1 Ph, 50 HZ) (110v, 1 ph, 60 HZ)
The machine is supplied with a 3 metre length of 1.5mm flex to BS6500 which
shall be connected to:
230 Volt Machine 16 Amp MCB to BS EN60898 Type C or fused 16 Amp to
IEC 60269.
110 Volt Machine 16 Amp MCB to BS EN60898 Type C or fused 16 Amp to
IEC 60269.
Notes:
1. All electrical installation to comply with current I.E.E. regulations.
2. Mains supplies to be protected by:
1 Phase 230v ......... 16 Amp approved I.E.C. fuse.
1 Phase 110v…….. 16 Amp approved I.E.C. fuse.
3. For all installations, an I.E.E. approved disconnection device must be fitted to the
final installation. This device shall be positioned in close proximity to the Vortex and
within easy reach of the operator & shall also be clearly marked as being the
disconnecting device for the machine.