6. Specifications
Classifications
Degree of protection against
electrical shock
Type BF
Protection against electric shock
Internally powered equipment
Electro-magnetic Compatibility
Group I, Class B
Apply part
Cuff and ECG electrode
Environmental
Item
Operating
Storage
Temperature
5
°
C to 45
°
C
-25
°
C to 70
°
C
Relative humidity (non-condensing)
10% to 95%
10% to 95%
Barometric
700hPa to 1060 hPa
700hPa to 1060 hPa
Degree of dust & water resistance
IP22
Drop test
1.0 m
Physical
Size
135 mm
×
46mm
×
21mm (main unit)
Weight
about 220 g (main unit)
Cuff size
22-42 cm
Wireless connectivity
Built-in Bluetooth 4.0 BLE
Power Supply
Charge input
Micro USB, DC5V
Battery type
Rechargeable lithium-polymer battery
Charge time
2 hours
Blood Pressure
Technology
Oscillometric method
Pressure measurement range
0
300 mmHg
Pressure measurement accuracy
±3mmHg or 2%, whichever is greater.
Pulse rate range
40 to 200 /min
Pulse rate accuracy
±2 /min
Clinical accuracy
Meet ISO
81060-2:2013
ECG
Lead type
Integrated ECG electrodes
ECG channel
Single Channel
Measure mode
Continuous
Input impendence
10MΩ, 10Hz