Victor VC2003 - Common Mode Operation; Inner Modulation Configuration; Outer Modulation Configuration; Linearity Frequency Sweep Setup

Victor VC2003
6 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
Press F key to set carrier frequency, input 6 , 3 , 0 , KHz/mV in turn to set output frequency
Press f
1
key to set the modulation signal as 1000Hz
Press % key and input 3 , 0 , %/S to set modulation depth
Press V key to set output amplitude, press 2 , MHz/V in turn to set output amplitude, and it completes.
During operation, the sequence is random to set carrier frequency, amplitude, depth and modulation signal.
4-2-3-3.
4-2-3-3.
4-2-3-3.
4-2-3-3.
OUTER-MODULATION
OUTER-MODULATION
OUTER-MODULATION
OUTER-MODULATION
SETTING
SETTING
SETTING
SETTING
When the inner-modulation parameter can not meet the requirement, you can set the instrument to outer-modulation mode.
Press
MOD
key to select outer-modulation mode, a nd
AM
and EXT/LOG
indicator light, the instrument is in outer-modulation mode. Input the outer-modulation signal from EXT
terminal
Press
F
key can set carrier wave
frequency
Press
V
key can set the amplitude of the output signal
Due to the modulation depth relates to the magnitude of the outer-modulation signal, the inner- modulation de pth
setting will not take function. The basic setting is the same as that of inner-modulation, and it is unnecessary to set modulation depth and frequency of inner-modulation signal.
The seemly frequency of outer-modulation signal is between 100Hz —— 100KHz
the input impedance
for
EXT
terminal is 600 Ω
when input signal is 0dB
,
the modulatio n
depth is 50%; when input signal is +6dB
,
the modulation depth is 100% the max. is approx.. 150%.
4-2-3-4.
4-2-3-4.
4-2-3-4.
4-2-3-4.
LINEARITY
LINEARITY
LINEARITY
LINEARITY
FREQUENCY
FREQUENCY
FREQUENCY
FREQUENCY
SWEEP
SWEEP
SWEEP
SWEEP
Press MOD
key to select linearity sweep, and Sweep
indicator light, the instrument is in linearity frequency sweep mode, press f
0
key to select initial frequency, then, th e
screen displays 0 ******
press
f
1
key to set terminate frequency, the screen
displays
1 ******
press
f
key to set spacing frequency, the screen
displays
2 * *****
press
t
key to set sweep time interval
After finishing
setting
t
it begins to cycle sweep according to positive linearity. It is suitable for sweeping in a narrower frequency b and.
For example, if you want to sweep between 1KHz to 1.1Khz by the frequency step value of 1Hz
,
set the time interval as 1s:
press MOD key to select linearity sweep mode
press f
0
key and 1 , KHz/mV key to set initial frequency
press f
1
key and 1 , . , 1 , KHz/mV key to set terminate frequency;
press f key and 1
,
Hz/%/S key to set sweep spacing frequency
press t key and 1 , %/S key to set time interval and then, sweep begins.
4-2-3-5.
4-2-3-5.
4-2-3-5.
4-2-3-5.
LOGARITHM
LOGARITHM
LOGARITHM
LOGARITHM
FREQUENCY
FREQUENCY
FREQUENCY
FREQUENCY
SWEEP
SWEEP
SWEEP
SWEEP
Press
MOD
key to select logarithm frequency sweep,
and
Sweep
and
EXT/LOG
indicator light, the instrument is in logarithm frequency sweep mode
press
f
0
key to
set
initiate frequency
press
f
1
key to set terminate frequency
press
t
key to set time interval. After finishing
setting
t
,
sweep begins . The setting way and displaying are
t
he sam e as
that of linearity sweep, and it is unnecessary to set f . It is suitable for sweeping in a wider frequency band.
For example, if want to make logarithm sweep between 20Hz and 20KHz
,
time interval is 0.1
s,
take the following
operation s :
Press MOD key to select logarithm sweep mode
Press f
0
and press 2
,
0 , Hz/%/S key to set initial frequency
Press f
1
key and press 2 , 0 , KHz/mV key to set terminal frequency
Press t key and press 0 , . , 1 , %/S key to set time interval, then, sweep begins.
EXPLANATION FOR FREQUENCY SWEEP
No matter which mode the frequency sweep is in, press any key s of f
0
,
f
1
,
f
,
t
to stop sweeping,
press
%/S
key to pause, then, the screen displays the freque ncy
of
the
pausing point, and will output this value. When press %/S
key again, it will go on sweeping. If need to restart,
press
t
key and
then
%/S
key.
The sweep begins with finis hing
input the unit of time interval t
,
so, time interval should be set lastly.
For
For
For
For
digital
digital
digital
digital
input
input
input
input
mode,
mode,
mode,
mode,
if
if
if
if
the
the
the
the
digit
digit
digit
digit
before
before
before
before
decimal
decimal
decimal
decimal
point
point
point
point
is
is
is
is
0
0
0
0
,
,
,
,
the
the
the
the
0
0
0
0
can
can
can
can
not
not
not
not
be
be
be
be
omitted.
omitted.
omitted.
omitted.
For any frequency output, TTL level square wave is in synchronism with sine output, and it is unnecessary to switch ranges; After power off or reset, the setting data on each
ranges become the defau l t value (F=1KHz,
V=1V,
DC offset= ± 0.15
V,
AM modulation frequency=400Hz, AM modulation amplitude m=50%, sweep initial frequency f
0
=1KHz, sweep
terminal frequency f
1
=10KHz, sweep interval frequency f =1KHz, sweep interval time t =1s).
5.
5.
5.
5.
NOTES
NOTES
NOTES
NOTES
5-1. There should not exist DC and high voltage and strong radiance signal on the load of the generator to prevent from damaging the instrument.
5-2. When the instrument appears the problem, such as un-normal displaying, breaking down. etc. just press reset key or turn off the power, the problem will be solved.
5-3. If there is a lower DC in load circuit or DC component is unnecessary
,
you can joint-connect a capacitor with proper capacitance and withstand voltage to interclude DC.
5-4.It
is
better not to connect a oversize load on TTL level output
,
otherwise, it will lead to wave form distortion, long rising time.
5-5. Do not expose the instrument to extremes of temperature and humidity, dusty place and strong radiance and electromagnetic interference environment, and avoid violent
shock.