- 2 -
to make measurement.
五、METHOD TO MEASURE THE INSULATED RESISTANCE
六、SAFETY NOTES:
1When voltage measurement selection button does not be
pressed, it is possible to appear the high voltage at the
output voltage terminal.
2When measuring, first check the measurement voltage
selection and if the indication of the measurement
voltage on LCD is in accordance with the voltage
needed
3To ensure the safety of operation, the operation cannot be
done until the measured object removed from the power
supplied by the electricity grid system, and should be
discharged fully to testify that the measured object does
not have any danger of power.
4When measuring, it is not allowed to handle the terminal
of measurement to make sure the accuracy of reading
and personal safety.
5Keep the instrument away from the high temperature
place, and from the direct sunshine to avoid affecting
the life of the LCD.
6When low battery symbol of appears, please
replace the batteries in time. Take the batteries out
when storage to avoid the leakage of batteries to
damage the instrument.
7When unload, reading is normal, it won’t affect the
measurement.
8When measuring MΩ, if the reading is unstable, it is
likely caused by the interference of surrounding and
unsteadiness of the insulated materials, it will make the
reading stable to connect the “G” terminal to the shield
side of the measured object
9To ensure the safety of measurement and reduce the
interference, the test leads adopt silicone rubber
materials, please do not change the leads at will.
10When the external AC adaptor supplies the power, the
inside batteries are disconnected and the batteries cant
be charged at the time. Note: please select ( )
power supply method.
七、SETS OF INSTRUMENT
1Digital Insulation Tester 1pc
210A test leads 1set
35# battery1.5V)×6 6pcs
4Operation Manual 1copy
5Shoulder strap 1pc
八、FAULT ELIMINATION
If the instrument can’t work properly, the below
methods can help you to solve the general problems quickly.
If the fault still can’t be eliminated, please contact the
maintenance centers or distributors.
Fault
Solution
No display
Power off-please turn on the
power
Replace the batteries
symbol
appearance
Replace the batteries
Error value
Replace the batteries
The instruction manual is subject to change without notice.
The contents in the instruction manual are considered to be
correct, if the user find any errors or pretermissions, etc.,
please contact the manufacturer.
The manufacturer hereby will not be responsible for any
accident and damage caused by the improper operation.
The functions described in this instruction manual do not be
the reason for special usage.
601E-60BX-000A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Victor VC60B+ and is the answer not in the manual?

Victor VC60B+ Specifications

General IconGeneral
BrandVictor
ModelVC60B+
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals