-2-
3. Classifications
[ Classification under the provision of 93/42/EEC(MDD)] ClassⅠ
ACP-900 is classified into ClassⅠ system.
[Protection method against electric shock] ClassⅠ
ACP-900 is Classified into ClassⅠ instrument.
ClassⅠ instrument in which protection against electric shock does not rely
on basic insulation only, but which includes an additional safety precaution
with the provision of the connection of the instrument to the protective
earth conductor in fixed wiring of the installation in such a way that any
accessible conductive parts cannot become live in the event of a failure
of the basic insulation.
[Degree of protection against ingress of liquids] IPX0
ACP-900 is the ordinary instrument (enclosed instrument without
protection against ingress of liquids). Be careful not to splash water
on the instrument.
[Degree of protection against flammability]
ACP-900 is classified as an equipment not suitable to be used in a
Potentially flammable anesthetic mixture with air or with oxygen
or nitrous oxide.
Do not use near flammable materials.
[Mode of operation]
Continuous operation