15.0 INSPECTION OF FOAM CONCENTRATES
As per NFPA 16, inspection of foam concentrates:
Periodic inspection shall be made at least annually of foam concentrates and their containers for excessive sludging or
deterioration. Inspection shall include a qualitative test of the foam concentrate normally conducted by the manufac-
turer.
Eight ounce (8 oz.) samples of foam concentrate should be taken from the top concentrate Valve No.5 on the tank, and
from the bottom concentrate drain/fill Valve No.7. Foam samples may be sent to the following address:
Dyne Technologies
2357 Ventura Drive, Suite 108
Woodbury, MN 55125
Tel: 651/917-0644
Fax: 651/917-0646
E-mail: jmleedy@dynetechnologies.com
Dyne Technologies offers three types of foam testing:
1. Foam Concentrate Testing is used to determine whether your undiluted foam concentrate still meets manufacturer
specifications, or whether it has degraded in storage.
2. Foam Solution Testing is used to ensure that your foam system is proportioning correctly.
3. On-site Inspection is performed addition to our in-laboratory testing services, Dyne Technologies can arrange for on-
site testing, inspection and maintenance of your foam and sprinkler systems through one of our experienced part-
nering organizations.
Contact Dyne Technologies for more details.
NOTE: DISCARD CONCENTRATE (PER MANUFACTURERāS RECOMMENDATIONS) IF FOUND DILUTED.
CORRECT THE SOURCE OF DILUTION BEFORE REPLACING THE NEW CONCENTRATE INTO TANK. CONSULT
VIKING FOR PROCEDURES ON TESTING INTEGRITY OF BLADDER.
16.0 BLADDER INTEGRITY TEST
With the schematics for valve descriptions and the air test equipment for your reference, the air integrity test of the blad-
der is as follows:
1. Isolate the tank from the water system by closing the water supply valve #2 and the concentrate valve #1.
2. Attach a garden hose to water drain valve #6, open, and terminate to the drain.
3. Slowly open water vent valve #4 and concentrate vent valve #5.
Note: Continue with the following steps when the water has completely drained.
4. Perform concentrate level check. (Refer to page 15)
5. Thread the air test equipment (see enclosed schematic) to concentrate vent valve #5. Connect an air supply to
the air regulator.
Note: Important to use 3-lb pressure gauge!
6. Begin filling the bladder with air until pressure gauge reads 1 PSI maximum. Do not be alarmed at the air escap-
18