14
from 1,0 to 33,9°C: ±0,3°C from 33.8 to 93.1°F: ±1.8°F
from 34,0 to 35,9°C: ±0,3°C from 93.2 to 96.7°F: ±0,5°F
from 36,0 to 39,0°C: ±0,2°C
(2)
from 96.8 to 102.2°F: ±0,4°F
(2)
from 39,1 to 43,0°C: ±0,3°C from 102.3 to 109.4°F: ±0,5°F
from 43,1°C to 80,0°C: ±1,0°C from 109.5 to 176.0°F ±1.8°F
(3)
When the HOME button is used in rooms where the temperature is in the range 5,0-9,9°C (41.0-
49.9°F) or 40,1-45,0°C (104.1-113.0°F), accuracy and operating range are not guaranteed, and the
temperature value is displayed alternately with the message “Lo.5» or “Hi.4” respectively.
Ambient temperature (shown when in stand-by)
Measurement range: 5,0/45,0°C (41.0/113.0°F)
Accuracy: ±1.0°C (±1.8°F)
13.2 General specifications
- Power supply: 4 AAA (LR03) alkaline batteries - 1.5 V (included)
- Life of high quality batteries: up to 3 years or 30,000 readings (depending on use)
- Dimensions: 144 x 43,5 x 21,5 mm (5.67 x 1.71 x 0.85 inches) - including cap.
- Weight: 98 gr. (3.46 oz.) - batteries included
- Distance from the subject: calculated using an optical aiming system (approx. 6
cm/2.36 inches).
- Large, clearly visible, backlit display in 5 colors: light blue, green or violet, depending
on the button used (respectively, “face”, “home” or “Mem”); orange and blue depen
-
ding on the function (respectively, room temperature – par. 5 and MQCS – par. 8)
- Expected life: 10 years.
- Atmospheric pressure range of operating conditions: from 700 hPa to 1,060 hPa.
- Relative humidity range of operating conditions: from 15% to 93%, non-condensing.
- Protection degree IP22
- Keep in a clean, dry place, preferably at a temperature ranging between +16 and
+40°C (60.8 and 104°F). Store at a temperature not lower than -25°C (-13°F) or higher
than +70°C (158°F).
- Internally powered equipment; mode of operation: continuous.
- The VisioFocus LEDs emit low light radiation in compliance with EN 62471.
- Measuring system tested in hospitals, private clinics and medical oces. Clinical
accuracy, characteristics and procedures are available from the manufacturer on re
-
quest.